LS126C UV Light Meter
Special for 254nm low-pressure mercury lamp
Record UV radiation intensity and dosage
Monitor real-time radiation intensity through the mobile APP
Provide PC software
LS126C UV light meter is a professional instrument for measuring the intensity of UVC germicidal lamps. The UV radiometer is delivered with a 1-meter test hook in accordance with the 1-meter test requirements specified in the “WST 367-2012 Regulation of disinfection technique in healthcare settings” to ensure full compliance with the test standards. The LS126C UVC light meter has a Bluetooth data transmission function. After connecting with the mobile phone APP, it can monitor the intensity of the germicidal lamp in real time from a certain distance to avoid UV burns during the test.
Parameter
Parameter | LS126C UV Light Meter |
---|---|
Spectral response | 230nm-280nm, λp = 254nm |
Measuring range | 0 – 20000 μW / cm² |
Resolution | 0.1μW / cm² |
Relative indication error | ±8%(relative to NIM standard) |
Long wave response error | <60% |
Cosine characteristics (directional response) error | 10% |
Linearity error | ±1% |
Short-term instability | ±1%( 30min after startup) |
Fatigue characteristics | attenuation <2% |
Error of zero | ±0.1% of full scale |
Response time | <0.5 seconds |
Sampling speed | 2048 times/second |
Test hole diameter | ¢=10mm |
Power value storage interval | 1s/10s/60s optional |
Recording period | 10 min / 100 min / 600 min |
Bluetooth transmission distance | 50 meters (open space) |
Instrument size | 148mm ×76mm ×26mm (L*W*H) |
Probe size | Diameter 39mm × thickness 15mm |
Probe wire length | 1 meter |
Weight | About 300 g |
Display | 240*160 dot matrix LCD |
Battery | 4 AAA alkaline dry batteries |
Operating Temperature | 0° to 40°C, <85% RH |
LS129 UV Digital Probe
Monitor UV intensity and energy in real time
Support MODBUS communication protocol
Provide RS485 communication interface
Connecting to PLC/Computer for easy automation integration
Small size for installation
Probe Parameter
Depending on the UV spectrum range and application industry, Linshang has launched three different UV digital probes to choose from, namely LS129-UVALED, LS129-UVA and LS129-UVC. The following table shows the relevant parameters of the three digital probes.
Probe Model | LS129-UVALED | LS129-UVA | LS129-UVC |
---|---|---|---|
Spectral response | 340nm-420nm, calibrated at 395nm |
315nm-400nm, λp =365nm |
230nm-280nm, λp =254nm |
Power measurement range | 0-20000mW/cm2 | 0-2000mW/cm2 | 0-20000μW/cm2 |
Energy measurement range | 0 – 4 × 109 mJ/cm2 | 0 – 4 × 109 μJ/cm2 | |
Resolution | 1mW/cm2 | 0.1mW/cm2 | 1μW/cm2 |
Measurement accuracy | ±10 % | ||
Sampling speed | 2048times/s | ||
Data fresh cycle | 200ms | ||
Test light hole diameter | ¢10mm | ||
Communication line length | 1m, resistant to high temperature of 200℃ | ||
Detector line length | 0.5m, resistant to high temperature of 200℃ | Integrated | |
Communication interface | RS485 | ||
Communication protocol | MODBUS | ||
Supply voltage | DC 5V-24V 20mA | ||
Operation temperature | -25℃- 85℃ (Humidity less than 85%, no condensation) | ||
Controller size | 72.5*32*15.1(mm) | ||
Probe size | 35*15*14.5(mm) | 35*23*16.7(mm) |
Communication Parameters
1. Station No
The LS129 smart digital probe supports the standard MODBUS protocol. The default communication address is “1”. You can modify the probe communication address through the “Human Machine Debugging Tool” or the communication protocol. The address setting range is 1-247.
2. Baud rate
The default baud rate of the UV digital probe is 9600bps. The baud rate can be modified through the “Human Machine Debugging Tool” or communication protocol.
3. Protocol
Supports the standard MODBUS protocol. You can read the measurement data, set the communication address and baud rate through the communication protocol. For detailed communication commands, please refer to “LS129 Digital Probe Communication Protocol”.
4. Probe wiring
Wire color | Function |
---|---|
Black (with heat-shrinkabletube) | Shielded wire |
Red | Positive power(DC 5-24V) |
Black | Negative power |
Green | RS485- |
White | RS485+ |
Feature
Adopt RS485 communication interface, Support MODBUS communication protocol
This UV intensity probe adopts standard RS485 communication interface and MODBUS communication protocol. It is equipped with high temperature wire capable of withstanding temperature of 200 ℃, which can realize online monitoring in real time for a long time
Whole probe is exquisite and compact, Detector and controller is seperate
The LS129 UV digital probe has a very small size, and the detector and controller adopt a split design, which is very convenient for installation where measurement is needed. It is also possible to combine the detector and controller with screws for measurement
Ultra-wide power supply range
The probe is powered by a 5-24V ultra-wide DC power supply range, which is very convenient for use in different industries on spot
Free computer debugging software for communication protocol debugging.
There are two kinds of RS485 negative communication cable in green and blue, please refer to the actual color of the received goods
LS125 Multi-probe UV Light Meter Host
Support 9 different probes for UVA, UVB, UVC, etc.
Record UV power, UV energy and test duration together
Automatic statistics of maximum, minimum and average power
LS125 is a smart host that can support 9 different probes. It can intelligently identify different probes. This host is connected to the probe through an aviation plug. The main function of the host is to display test data. The host equipped with the probe can be used for ultraviolet intensity and energy detection. To use the host, you need to buy the corresponding probe to use with it as per your test requirements.
Parameter
Parameters | LS125 UV Light Meter |
---|---|
Weight | About 194g |
Dimension | 148mm × 76mm × 26mm (L*W*H) |
Display | 240*160 dot matrix LCD |
Power supply | 4 AAA alkaline batteries |
Probe connection | Push pull (Snap-in style),Aviation socket |
Application
Ultraviolet light includes UVA band (315nm-420nm), UVB band (280nm-315nm) and UVC band (230nm-280nm). UVA band is mainly used in photocuring, lithographic exposure, ultraviolet light source, integrated circuit lithography and other industries; UVB band UV lamps mainly include physiotherapy lamps in hospitals and health care places, plant growth lamps and UVB fluorescent lamps for reptiles; UVC band also known as short-wave sterilization ultraviolet, ultraviolet germicidal lamps emit UVC short-wave ultraviolet. Ultraviolet germicidal lamps are mainly used for disinfection and sterilization. They are often used in hospitals, transfer windows of drug factories, disease control centers and sewage treatment plants. The LS125 handheld UV measurement device is widely used. Many of the above industries require different ultraviolet probes to detect the radiation intensity or energy value.
Probe Selection
According to the spectral response range, application, measuring range and test aperture size of the light source, we have introduced 9 different types of probes (probe must be purchased separately). The selection table is as follows:
Select according to the spectral response range and application of the light source
No. | Probetype | Spectral response range | Power range | Test aperture | Application | Link |
---|---|---|---|---|---|---|
1 | UVC-X0 | 230-280nm | 0 – 200000 μW/cm² | ¢10mm | 254nm UV sterilization mercury lamp | More details |
2 | UVCWP-X1 | 230-280nm | 0 – 200000 μW/cm² | ¢10mm | 254nm UV sterilization mercury lamp,waterproof | More details |
3 | UVCLED-X0 | 240-320nm | 0 – 200000 μW/cm² | ¢10mm | 260nm-285nm UV LED germicidal lamp | More details |
4 | UVB-X0 | 280-315nm | 0 – 200000 μW/cm² | ¢10mm | General UVB intensity and energy measurement | More details |
5 | UVA-X0 | 315-400nm | 0 – 2000 mW/cm² | ¢10mm | High pressure mercury lamp in UV curing industry | More details |
6 | UVA-X1 | 315-400nm | 0 – 200000 μW/cm² | ¢10mm | General UVA intensity and energy measurement | More details |
7 | UVALED-X0 | 340-420nm | 0 – 20000 mW/cm² | ¢10mm | Area light source of UV LED in UV curing industry | More details |
8 | UVALED-X1 | 340-420nm | 0 – 20000 mW/cm² | ¢1mm | Point light source of UV LED in UV curing industry | More details |
9 | UVALED-X3 | 340-420nm | 0 – 200000 μW/cm² | ¢10mm | General UVA+UVV LED, low power measuring range | More details |
Suggestion of Probes for Classic Industry Application
No. | Application | Suggested Probe |
---|---|---|
1 | Detect UV aging box | UVA-X1、UVB-X0 |
2 | Detect ambient light and sunlight |
UVA-X1、UVB-X0 |
3 | Detect welding arc | UVA-X1、UVB-X0、UVC-X0 |
4 | Detect UV light of non-destructive testing |
UVA-X0 |
5 | Detect UV Nail Lamp | UVALED-X3 |
6 | Detect plant growth lamp or reptile lamp |
UVB-X0 |
Free to match 9 types of probes
LS125 host and probe are linked by aviation plugs. One host can intelligently match 9 different UV probes to meet different test needs
LS125 UV Light Meter +UVCWP-X1 Probe
Suitable for the testing of high-power germicidal lamp in water treatment industry
Spectral response: 230nm-280nm, λp = 254nm
Waterproof, the probe can be submerged to the depth of 1 meter
Large power range , the maximum measurement power up to 200000μW/cm²
The LS125 UV light meter is a smart host that can support 9 different types of probes. Its main function is data display. The UVCWP-X1 probe is one of the 9 probes equipped with the LS125 host. The probe’s spectral response range is 230nm-280nm, the peak wavelength is 254nm. This UV radiometer is waterproof to 1 meter deep. It is a probe specially developed for water treatment high-power germicidal lamps.
Application
Water treatment is divided into wastewater treatment, urban water supply treatment, drinking water treatment, purified water treatment, and agricultural wastewater treatment. There are two methods of water treatment. The first is to put sterilization lamps into the water to sterilize. The second method is to sterilize the lamp in a sleeve. Because the water flow rate is too fast, the germicidal lamp time is sometimes less than 1 second, so a high-power germicidal lamp or a longer ultraviolet lamp is required to sterilize. LS125 + UVCWP-X1 waterproof light meter is a professional instrument for detecting water treatment germicidal lamps.
If you want to know about other types of probes supported by the LS125 UV light meter host, please click “LS125 Multi-Probe UV light meter Host” for details
Parameters
Parameters | UVCWP-X1 Probe |
---|---|
Spectral response | 230nm-280nm, λp = 254nm |
Power measuring range | 0 – 200000μW / cm² |
Resolution | 0.1μW / cm² |
Energy measuring range | 0-9999999μJ/cm² |
Record time | 0-99999s |
Measuring accuracy | ±10% |
Sampling speed | 6 times/second |
Unit | μW / cm² (default), mW / cm², W / m² |
Test Hole Diameter | ¢=10mm |
Probe size | Diameter 50mm × thickness 20mm |
Waterproof depth | 1m |
Packing box dimension | 24 * 6 * 5 cm (L * W * H) |
Large power range , the maximum measurement power can reach 200000μW/cm²
Large-scale measurement, the maximum power can be measured to 200,000 μW / cm²
Data Statistics
Units are available
Three measurement units can be switched, namely μW/cm², mW / cm², W / m². Data conversion between different units does not require manual calculation
Qualifications and Standards
The UV light meter LS125 host + UVCWP-X1 probe conform to the following standard:
JJG 879-2015 Verification Regulation of Ultraviolet Radiometers
Reviews
There are no reviews yet.